Hanna HI98198 measures dissolved oxygen levels in water using enhanced optical dissolved oxygen technology to ensure high levels of accuracy. With built-in methods and calculations for the measurement of Biological Oxygen Demand (BOD), Oxygen Uptake Rate (OUR), and Specific Oxygen Uptake Rate (SOUR), there is no need for manual conversions.
Increased Traceabilty
Ensure that your data can be trusted with GLP data. Accessible by a fixed dedicated key, this data includes date, time, pressure, calibrated value, temperature and salinity value of the last calibration.
Versatile Data Logging
The HI98198’s log-on-demand feature allows you to store up to 4000 readings. The AutoEnd button can be used during measurement to hold the first stable reading on the display automatically.
Focus on What Matters
When you put on a new cap, your meter will detect and initialize it. It will remind you when your cap needs to be replaced, is damaged, or is missing, giving you one less thing to worry about.
Specifications Hanna HI98198 :
- OXYGEN, DISSOLVED MEASUREMENT TYPE : Optical – luminescence
- OXYGEN, DISSOLVED RANGE : 0.00 to 50.00 mg/L (ppm); 0.0 to 500.0% saturation
- OXYGEN, DISSOLVED RESOLUTION : 0.01 mg/L (ppm); 0.1% saturation
- OXYGEN, DISSOLVED ACCURACY : 1.5% of reading ± 0.01mg/L for 0.00-20.00mg/L , 5% of reading for 20.00-50.00mg/L , 1.5% of reading ±0.1% for 00-200.0%, 5% of reading for 200.0-500.0%
- OXYGEN, DISSOLVED CALIBRATION : one or two points automatic calibration at 100% (8.26 mg/L) and 0% (0 mg/L); Single point manual using a value entered by the user in % saturation or mg/L.
- OXYGEN, DISSOLVED TEMPERATURE COMPENSATIO : Nautomatic from -5.0 to 50.0 °C (23.0 to 122.0 °F)
- BAROMETRIC PRESSURE COMPENSATION : automatic from 420 to 850 mmHg
- SALINITY COMPENSATION: automatic from 0 to 70 g/L (manually set)
- ATMOSPHERIC PRESSURE RANGE : 420 to 850 mmHg (inHg, atm, mbar, psi, kPa units available)
- ATMOSPHERIC PRESSURE RESOLUTION : 1 mm Hg
- ATMOSPHERIC PRESSURE ACCURACY : ± 3 mmHg within ±15% from the calibration point
- ATMOSPHERIC PRESSURE CALIBRATION: one point at anywhere in pressure range value
- TEMPERATURE RANGE : -5.0 to 50.0 °C (23 to 122 °F)
- TEMPERATURE RESOLUTION : 0.1 °C (0.1 °F)
- TEMPERATURE ACCURACY : ±0.3 °C (±0.4 °F)
- TEMPERATURE CALIBRATION : Single point anywhere within temperature range
- TEMPERATURE COMPENSATION : Automatic
- DO PROBE : HI764113 optical DO probe with protective sleeve, internal temperature sensor, quick connect DIN connector and 4m (13’) cable (included)
- LOGGING TYPE : On demand with 4000 records capability
- CONNECTIVITYUSB-CAUTOMATIC SHUT-OFF: User selectable: 5, 10, 30, 60 minutes or disabledGLPYes- includes last 5 calibrations
- CALIBRATION TIMEOUT REMINDER : Yes- user selectable “Calibration due” warning
- AUTOEND : Yes
- BOD/OUR/SOUR : Yes
- DISPLAY : LCD (back lighted display)
- BATTERY TYPE/LIFE : 1.5V AA batteries (4 pcs.) / approx. 200 hours of continuous use without backlight (50 hours with backlight)
- IP PROTECTION : IP67
- ENVIRONMENT : 0 to 50 °C (32 to 122 °F) max. RH 100%
- DIMENSIONS : 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
- WEIGHT : 450 g (15.9 oz.) with batteries
- ORDERING INFORMATION : HI98198 is supplied with HI764113 optical DO probe, smartcap with RFID, HI7040 bicomponent zero oxygen solution, 100 mL plastic beakers (2), HI920016 micro USB Type A to C cable, 1.5V AA batteries (4), instruction manual, quick start guide, instrument quality certificate and a rugged carrying case.